Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device received with a lot of labels on it.Visual inspection noted the quick drain connect was corroded, water tank cover had multiple cracks where the screws were put in, plate clip had a nick in it, front cover had multiple scratches, pole clamp had deep gouges in it, water tank was contaminated.When performing the functional tests, the pump did not turn on.The complaint was confirmed.The root cause was due to an inoperative water pump and no longer recirculating water.It was unknown what caused the pump to fail.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the pump, quick connect, water tank cover, plate clip, pole clamp and front cover.Performed preventative maintenance (pm) and calibration.Device passed all functional and delivery tests.
|