U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Overheating of Device (1437); Failure to Pump (1502)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Event Description
It was reported that the device was not pumping water and was overheating.Patient involvement is unknown.
 
Event Description
Additional information received via email and attached to complaint object: the issue was found during a routine preventative maintenance (pm).No patient was involved.
 
Manufacturer Narrative
Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device received with a lot of labels on it.Visual inspection noted the quick drain connect was corroded, water tank cover had multiple cracks where the screws were put in, plate clip had a nick in it, front cover had multiple scratches, pole clamp had deep gouges in it, water tank was contaminated.When performing the functional tests, the pump did not turn on.The complaint was confirmed.The root cause was due to an inoperative water pump and no longer recirculating water.It was unknown what caused the pump to fail.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the pump, quick connect, water tank cover, plate clip, pole clamp and front cover.Performed preventative maintenance (pm) and calibration.Device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16877535
MDR Text Key314931383
Report Number3012307300-2023-05162
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/05/2023
Supplement Dates Manufacturer Received05/08/2023
08/24/2023
Supplement Dates FDA Received05/26/2023
08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-