RESPIRONICS, INC. DREAMSTATION 2 ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX520H11C
Device Problems Complete Loss of Power (4015); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2023
Event Type  malfunction  
Event Description
The manufacturer received information alleging the dreamstation 2 advanced auto cpap device would not power on and they tried to use other cords.There was no report of patient harm or injury.The device was returned to the manufacturer for investigation.During the evaluation of the device the technician noted a failed pca due to evidence of water on the pca.Secondary findings noted thermal damage to q2.The technician was unable to retrieve the log data.The unit is being sent to the product investigation laboratory for further evaluation.The unit is being sent to the product investigation laboratory for further evaluation.
 
Manufacturer Narrative
Corrective and preventative actions have been opened by the manufacturer to address the water ingress identified upon investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2 ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16878044
MDR Text Key314660244
Report Number2518422-2023-10534
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520H11C
Device Catalogue NumberDSX520H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No