Model Number HL-390 |
Device Problems
Insufficient Heating (1287); Failure to Calibrate (2440)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: date of event is unknown, no information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the warmer would not heat or calibrate.Patient involvement unknown.
|
|
Event Description
|
Additional information received via email.The event occurred during set up.No patient was involved.
|
|
Manufacturer Narrative
|
Other text: h6 evaluation codes: updated.One device was returned for investigation with a little dirt on front cover, top and back by the pole clamp.Reflux plug is missing and water tank has some rust stains.Visual and functional testing was conducted on the device and the reported problem could not be duplicated.However, it was found that the device was over heating.It is possible that a faulty pump not circulating water caused this issue.After replacing the pump the device passed all functional tests.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.
|
|