OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the warmer displayed over temperature warning at start up.No patient involvement was reported by the customer.
 
Manufacturer Narrative
Evaluation codes: updated device evaluation: one device was returned for investigation.Visual inspection noted the device was received in very good condition.Functional testing could not duplicate or confirm the reported problem.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.Preventative maintenance completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16889654
MDR Text Key314772550
Report Number3012307300-2023-05318
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-