Model Number HL-90
Device Problems Device Alarm System (1012); Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the unit was leaking and over temperature alarm was not working properly.Patient involvement unknown.
 
Manufacturer Narrative
Date of event and udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One device was received.Visual inspection: corroded drain fitting, cracked tank cover, and broken liquid crystal display (lcd).Functional test: filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The device leaked from a corner of the reservoir and the over temperature alarm did not sound when it should, confirming the complaint.Cracks in the tank cover caused the leaking and the pot used to set the over temperature alarm was damaged however it was unknown what caused the damage.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Event Description
Additional information received: the event date is unknown.There was no patient harm or injury.
 
Manufacturer Narrative
Other text: b5, h6 health effects and event problem device code: updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16890094
MDR Text Key314826376
Report Number3012307300-2023-05325
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-