Model Number HL-90
Device Problems Device Alarm System (1012); Display or Visual Feedback Problem (1184)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the over temperature alarm and "lad" was not working.Patient involvement unknown.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received dirty, front cover broken, water tank cracked, float switch was stained, microswitch had broken lever, serial number and "tuv" were torn.Functional testing confirmed the complaint as the device failed on water level; when depressing the float switch the unit did not alarm, and led did not light.The root cause was due to a faulty printed circuit board (pcb).It was unknown what caused the fault.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced printed circuit board (pcb), float switch, water tank cover, microswitch, line cord, front cover.Performed preventative maintenance (pm) and calibrated.Device passed all functional and delivery tests.
 
Event Description
Additional information was received confirming the issue with this device was discovered during a routine preventive maintenance procedure.No known patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16890890
MDR Text Key314780521
Report Number3012307300-2023-05341
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2023
Initial Date FDA Received05/08/2023
Supplement Dates Manufacturer Received08/29/2023
09/05/2023
Supplement Dates FDA Received08/30/2023
09/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-