• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2023
Event Type  Injury  
Event Description
Related manufacturer reference number (b)(4).It was reported that patient experienced ineffective therapy.System diagnostics recorded high impedance on one lead.Surgical intervention was undertaken on (b)(6) 2023, wherein the surgeon was unable to remove both leads due to scar tissue.The right l4 lead was broken while the left l4 lead was cut.(related manufacturer reference number (b)(4)).
 
Manufacturer Narrative
The event of fragment left behind was reported to abbott.The patient was explanted due to insufficient relief and inability to get mri (covered on per-2023-03112).During the case, the surgeon was unable to remove both leads due to scar tissue.Right l4 lead broke upon removal attempt.Surgeon cut the tail on the left l4 lead due to tension when trying to remove.The results of the investigation are inconclusive as the device was not returned for evaluation.Based on the information received, a single definitive root cause for the issue encountered was unable to be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key16892367
MDR Text Key314808368
Report Number1627487-2023-02152
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2023
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number8009633
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/25/2023
Initial Date FDA Received05/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DRG IPG.
Patient Outcome(s) Other;
Patient Weight91 KG
-
-