Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer's allegation was confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus the 10mm rigid telescope, hd, quick lock had defective optics.The issue was found during preparation for use in an unidentified, therapeutic procedure which was completed using a similar device.Inspection and testing of the returned device found the device eyepiece was damaged.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/h10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the reported event occurred due to the use of excessive force from improper handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID OPTICAL LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16896993
MDR Text Key314847169
Report Number9610773-2023-01259
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Catalogue NumberWA53005A
Device Lot Number697645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received07/24/2023
Supplement Dates FDA Received08/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-