SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 21-7346-24 |
Device Problem
Inaccurate Delivery (2339)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the disposable caused an under infusion for 2 days after tubing changed.No adverse patient effects were reported by the customer.
|
|
Manufacturer Narrative
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: one sample was received in used condition.During the visual inspection, no damage was found on the sample.During the functional testing, no discrepancies were found on the sample.The test successfully passed with results within spec.The customer reported failure was unable to be confirmed.A root cause was unable to be determined.No relevant non-conformities or findings were found during the dhr review.
|
|