MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7346-24
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Event Description
It was reported the disposable caused an under infusion for 2 days after tubing changed.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Device evaluation: one sample was received in used condition.During the visual inspection, no damage was found on the sample.During the functional testing, no discrepancies were found on the sample.The test successfully passed with results within spec.The customer reported failure was unable to be confirmed.A root cause was unable to be determined.No relevant non-conformities or findings were found during the dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SETS WITH FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16897049
MDR Text Key314850570
Report Number3012307300-2023-05408
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Device Lot Number4291448
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received07/22/2023
Supplement Dates FDA Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-