Catalog Number 71331956
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Joint Dislocation (2374)
Event Date 04/14/2023
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that after a thr surgery had been performed on (b)(6) 2023 of an r3 3 hole ha ctd acet shell 56mm, the patient experienced a dislocation.This adverse event was solved by revision surgery on (b)(6) 2023.The current health status of the patient is unknown.
 
Manufacturer Narrative
D11: concomitant medical products and therapy dates and g5: pma/510(k) #.Section h3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that there were no clinical factors found which would have contributed to the reported dislocation.The patient's current condition is unknown and the patient impact beyond the revision surgery could not be determined.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that dislocation may result from improper selection of the neck length and cup, stem positioning, muscle looseness and/or mispositioning of components.This has been identified in warnings and precautions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient anatomy, postoperative care or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D2b: procode and h6: health effect - clinical code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE HA CTD ACET SHELL 56MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16897138
MDR Text Key314848496
Report Number1020279-2023-01010
Device Sequence Number1
Product Code MBL
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K201701
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71331956
Device Lot Number22EM05243
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received05/30/2023
Supplement Dates FDA Received06/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Treatment
OXINIUM FEM HD 12/14 36 MM M/+4, LOT#:22HM30328; R3 20 DEG XLPE ACET LNR 36MM X 56MM, LOT#:22HM1564
Patient Outcome(s) Required Intervention;
-
-