• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FEMORAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FEMORAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS Back to Search Results
Model Number 2544-01-006
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.The visual analysis of the returned sample revealed that the attune femoral impactor had several cracks across the surface.Additionally, the device was broken, the broken fragments were not returned for examination.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Instruments were damaged in sterilization.It did not happen in surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEMORAL IMPACTOR
Type of Device
ATTUNE INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16897984
MDR Text Key314860723
Report Number1818910-2023-09946
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295130222
UDI-Public10603295130222
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2544-01-006
Device Catalogue Number254401006
Device Lot NumberAU6267280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received05/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-