Model Number 0475000000
Device Problem Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the procedure was cancelled.The patient was under anesthesia, but no incision was made.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device did not boot up.Probable root cause: software error.Front board issue.Blown fuse.Power board issue.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.Insufficient cybersecurity (cf).The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Event Description
It was reported that the procedure was cancelled.The patient was under anesthesia, but no incision was made.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSFIRE CONSOLE INTERNATIONAL KIT
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16898803
MDR Text Key314887194
Report Number0002936485-2023-00405
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327058086
UDI-Public07613327058086
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0475000000
Device Catalogue Number0475000000I
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received04/10/2023
Supplement Dates FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-