• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7047-24
Device Problems Excess Flow or Over-Infusion (1311); Product Quality Problem (1506); Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the disposable caused the device to alarm and deliver infusion two hours early.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
Other text: d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one sample was received.The visual inspection found no damage.During the functional testing, the sample was connected to a filled cassette and a pump.The pump was set to run and no alarms were activated.The customer reported failure was not confirmed.A root cause was unable to be confirmed.There was no non-conformance or deviation related to the failure reported in the device history record for finish good and component associated.By the date of this complaint investigation report no complaint related to the failure mode and lot number reported in the last 12 months.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16902840
MDR Text Key314912818
Report Number3012307300-2023-05466
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586020315
UDI-Public10610586020315
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K942046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7047-24
Device Catalogue Number21-7047-24
Device Lot Number4354707
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2023
Initial Date FDA Received05/10/2023
Supplement Dates Manufacturer Received07/22/2023
Supplement Dates FDA Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/03/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-