SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS WITH FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 21-7346-24 |
Device Problem
Inaccurate Delivery (2339)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/11/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported the disposable under delivered.Client has a daily continuous dose of iv antibiotics that is administered via tunneled cvc.Dose runs over 30 minutes.Medication was primed through tubing without issues.At the end of administration approx.10-20mls of medication remained in bag.Typically, bag will be empty by the end of the dose.No adverse patient effects were reported by the customer".
|
|
Manufacturer Narrative
|
No product was returned.The reported issue could not be confirmed.If the product is returned this complaint will be reopened for further investigation.No lot number was provided; therefore, a device history record (dhr) could not be conducted.
|
|