Other text: d4, d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one sample was received in unused condition.During the visual inspection, no damage was found.During the functional testing, the cassette was set up to be used with a pump.The fluid was able to flow through the whole device and no alarms were activated.The customer reported fault was unable to be confirmed.A root cause was unable to be confirmed.There were no relevant findings detected in the dhr.
|