Model Number HL-390 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the unit was leaking around the front plastic clear casing.Patient involvement unknown.
|
|
Manufacturer Narrative
|
Device evaluation: one device was returned for evaluation.The tank was severely cracked, drain fitting corroded, block assembly worn and a faded line cord.The tank was filled with water and temp check was attached and power switch turned on.The device leaked from the reservoir confirming the customer complaint.No action was taken as the device was deemed beyond economical repair.Device was scrapped.Manufacturing device history review was not done due to the device being beyond a year from the manufacturing date.Service history review showed device had not been in for service previously.
|
|