Catalog Number 364327
Device Problem Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2023
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd preset¿ abg syringe with luer-lok tip and attached needle there was missing scale label information.The following information was provided by the initial reporter.The customer stated: stage: before ready to use.Defect: it is found that the product has no scale.Quantity: 1 piece.Effects: no adverse effects.
 
Manufacturer Narrative
The following fields have been updated with additional information: d.10 device available for eval? yes.D.10 returned to manufacturer on: 2-may-2023.Investigation summary: bd received 1 sample for investigation.The sample was evaluated by visual examination and the indicated failure mode for missing scale markings with the incident lot was observed.Additionally, 10 retention samples from bd inventory were evaluated by visual examination and the issue of missing scale markings was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode missing scale markings.Bd was not able to identify a root cause for the indicated failure mode.
 
Event Description
It was reported when using the bd preset¿ abg syringe with luer-lok tip and attached needle there was missing scale label information.The following information was provided by the initial reporter.The customer stated: stage: before ready to use.Defect: it is found that the product has no scale.Quantity: 1 piece.Effects: no adverse effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRESET¿ ABG SYRINGE WITH LUER-LOK TIP AND ATTACHED NEEDLE
Type of Device
SYRINGE WITH NEEDLE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16908354
MDR Text Key315020737
Report Number9617032-2023-00537
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K022426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2024
Device Catalogue Number364327
Device Lot Number2091539
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-