|
Model Number 53401 |
Device Problems
Operating System Becomes Nonfunctional (2996); Output Problem (3005)
|
Patient Problems
Asystole (4442); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the external pulse generator (epg) stopped pacing while in use. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product analysis: analysis was unable to confirm the customer comment that the epg provided only intermittent pacing or stopped pacing.It was noted that upon performance of check according to procedure it was found to work as expected and was ready for use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that a patient was connected to the external pulse generator (epg) used with patient cable when the epg provided only intermittent pacing or stopped pacing.The patient was dependent on the epg and required cardiopulmonary resuscitation (cpr).It was speculated by the customer that the issue may have been related to the position of the patient cable when connected to the epg.The epg and patient cable were replaced.The patient was hospitalized in the intensive care unit (icu) as a result of this event.
|
|
Search Alerts/Recalls
|
|
|