• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500T11C
Device Problems Contamination (1120); Degraded (1153)
Patient Problems Dyspnea (1816); Headache (1880); Dizziness (2194); Sneezing (2251); Cough (4457)
Event Date 07/06/2021
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged of having dizzy, coughing, sneezing, headaches and breathing problem.There was no report of serious patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection of the device was completed by the manufacturer and found evidence of liquid ingress and an unknown dust on the blower, blower box and bottom enclosure.The manufacturer also found an unknown contaminant and a brown congealed substance on the top enclosure, modem area, sd card slot, front panel and on the bottom enclosure.The manufacturer found no evidence of sound abatement foam degradation/breakdown.The device's downloaded event log was reviewed by the manufacturer and found no errors logged.The device was applied power and the device operated properly.The manufacturer concludes the contaminates found were consistent with liquid ingress, an unknown contaminant, dust and brown substances, contamination inconsistent with the sound abatement foam.The manufacturer confirmed there was no evidence of sound abatement foam degradation/breakdown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16917238
MDR Text Key315073105
Report Number2518422-2023-11008
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2023
Date Manufacturer Received04/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-