• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Event Description
During the initial implant procedure, the stylet was unable to be inserted into the left ventricular (lv) lead.The lv lead was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of failure to advance stylet/ guidewire was confirmed.As received, a complete lead was received in one piece.The stylet/ guidewire used in the field was not returned with the lead.X-ray examination of the lead found the inner coil inside the connector region was bunched up consistent with procedural damage.Unable to perform stylet/ guidewire insertion test due to the bunched up inner coil.The cause of the reported event was isolated to the bunched up inner coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16920291
MDR Text Key315107759
Report Number2017865-2023-19825
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberS000089346
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/27/2023
Initial Date FDA Received05/12/2023
Supplement Dates Manufacturer Received06/02/2023
Supplement Dates FDA Received06/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-