• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL 8CH INFINITY DBS FLEX EXTN KIT, 60CM, B; DBS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL 8CH INFINITY DBS FLEX EXTN KIT, 60CM, B; DBS EXTENSION Back to Search Results
Model Number 6372
Device Problem Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/27/2023
Event Type  Injury  
Event Description
Related manufacturer reference number 1627487-2023-01304.It was reported the patient experienced ineffective therapy.Diagnostics indicated that the leads had multiple contacts with low and high impedance.Surgical intervention was undertaken where in the extension was explanted and replaced restoring therapy.
 
Manufacturer Narrative
B3-date of event is estimated.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CH INFINITY DBS FLEX EXTN KIT, 60CM, B
Type of Device
DBS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key16921609
MDR Text Key315118301
Report Number1627487-2023-02153
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067030351
UDI-Public05415067030351
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/15/2021
Device Model Number6372
Device Catalogue Number6372
Device Lot Number6800156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DBS BURR HOLE CAP (X2); DBS EXTENSION; DBS LEAD (X2)
Patient Outcome(s) Other;
Patient SexMale
Patient Weight93 KG
-
-