MEDTRONIC, INC. MEDTRONIC LEAD; IMPLANTABLE PACEMAKER PULSE-GENERATOR Back to Search Results
Model Number MDT-LEAD
Device Problems High impedance (1291); Over-Sensing (1438); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  Injury  
Manufacturer Narrative
This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.Of note, multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique product serial/lot numbers.The model listed in the report is a representative of the model family, as there is no specific model listed.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: incidence and predictors of cardiac implantable electronic devices malfunction with radiotherapy treatment.Journal of clinical medicine.2022 oct 27;11(21):6329.Pmid: 36362559 doi: 10.3390/jcm11216329 digital sharing of the article would be in violation of copyright permission.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Literature was reviewed regarding cardiovascular implantable electronic device (cied) malfunction with radiotherapy (rt) treatment.The authors described patients who underwent rt treatment in the radiation oncology unit.There were implantable cardioverter defibrillators (icds) which exhibited partial electrical resets in which the devices were reprogrammed and unexpected decline in battery four months post rt with an urgent replacement.There was an implantable pulse generator (ipg) which exhibited an unexpected decline in battery two months post rt and it was explanted and replaced.There were right ventricular (rv) leads which exhibited increased impedance, undersensing, and oversensing with reprogramming.The status of the devices and leads is unknown.No adverse patient effects or additional product performance issues were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC LEAD
Type of Device
IMPLANTABLE PACEMAKER PULSE-GENERATOR
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16921836
MDR Text Key315120334
Report Number2182208-2023-01359
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature
Reporter Occupation Physician
Type of Report Initial
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMDT-LEAD
Device Catalogue NumberMDT-LEAD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2023
Initial Date FDA Received05/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN COMPETITOR ICD/IPG
Patient Outcome(s) Required Intervention;
-
-