RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V200
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the esprit indicating that the device has screen issues, when turning on after a few minutes screen goes blurry and gives lines.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The biomed customer was tapping on the side of the unit and it seemed to solve the issue.The rse recommended him to open the gui assembly, and check for loose cables.Also make sure the transition pcb is properly connected and the display cable is secure.Investigation is ongoing.
 
Manufacturer Narrative
H10: the customer opened and checked for loose cables.Nothing was loose or disconnected.He switched on and off about 10 times, no issues found or duplicated.Ran the unit in normal operation mode for 3 hours, no issues found or duplicated.Performance procedure tests passed.Customer decided to remove unit from service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key16921890
MDR Text Key315136647
Report Number2518422-2023-11071
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV200
Device Catalogue Number1060264
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/09/2023
Date Device Manufactured03/18/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse