Model Number CI-1500-01 |
Device Problems
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Electro-Static Discharge (2149)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 05/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
A review of the recipient¿s test data indicated impedance issues.Device testing revealed results within normal limits.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The failure of this device is attributed to a short from a power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
A review of the device history record was completed and no anomalies were noted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
A ct scan revealed correct device placement.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|