Model Number 1458Q/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  malfunction  
Event Description
It was reported that during an implant procedure that the left ventricular (lv) lead was unable to be advanced smoothly in the guidewire.When attempting to implant a different lv lead the guidewire broke.The physician elected to complete the procedure with a left bundle branch pacing lv lead.The patient condition was stable.
 
Manufacturer Narrative
The reported event of the guidewire could not be inserted was confirmed.As received, a complete lead was returned in one piece.X-ray examination of the lead found the inner coil bunching up at the connector region.The cause of the reported event was isolated to the inner coil damaged consistent with procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16924438
MDR Text Key315275227
Report Number2017865-2023-19858
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000129613
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/12/2023
Supplement Dates Manufacturer Received06/08/2023
Supplement Dates FDA Received06/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-