• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; MECHANICAL ACC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; MECHANICAL ACC Back to Search Results
Model Number A22040A
Device Problems Break (1069); Crack (1135); Component Missing (2306); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and repair.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.Additional pma/510(k): k931995.
 
Event Description
It was reported to olympus that an inner sheath was dropped after surgery.The tip peak part was missing, and the device was cracked.There was no patient injury or adverse event reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Sections h3, h6, and h10 were updated.Section g2 was corrected with information that was inadvertently omitted from the initial report.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the damaged ceramic tip could not be determined.It is possible that the damage was induced thermally or mechanically, such as through fall, shock, or similar stress.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿4 before use: warning: infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing: inspecting the product: visually inspect the product.Make sure that it has: -- no corrosion., -- no dents, and -- no scratches.Ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury: impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
MECHANICAL ACC
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16928182
MDR Text Key315274581
Report Number9610773-2023-01316
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761029339
UDI-Public04042761029339
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberA22040A
Device Catalogue NumberA22040A
Device Lot Number09YW-0051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/13/2023
Supplement Dates Manufacturer Received08/14/2023
Supplement Dates FDA Received08/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-