MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number FINETUNER ECHO [FINETUNER 2]
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
A finetuner echo was returned to service and repair due to mechanical problems.
 
Event Description
A finetuner echo was returned to service and repair due to mechanical problems.
 
Manufacturer Narrative
Conclusion: according the information received, a finetuner echo was sent to service _ repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16929643
MDR Text Key315185522
Report Number9710014-2023-00402
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737344290
UDI-Public(01)09008737344290
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberFINETUNER ECHO [FINETUNER 2]
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/08/2023
Initial Date FDA Received05/15/2023
Supplement Dates Manufacturer Received05/08/2023
Supplement Dates FDA Received07/11/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-