MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US EA DELTA CER INSERT 36IDX52OD; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US EA DELTA CER INSERT 36IDX52OD; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS Back to Search Results
Catalog Number 121881752
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 01/01/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient had postoperative infection and was hospitalized for observation.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicates that a debridement surgery was performed on (b)(6) 2023, and now the patient was still in the hospital for observation.
 
Manufacturer Narrative
Product complaint # (b)(4) investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that the device was broken.The broken fragment was not visible in the provided evidence.The breakage of the device was captured in pi-1680228851541087.Based on the provided evidence, there is no indication that the complaint device contributed to reported adverse event.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EA DELTA CER INSERT 36IDX52OD
Type of Device
PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16930542
MDR Text Key315202350
Report Number1818910-2023-10255
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number121881752
Device Lot Number3785510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2023
Initial Date FDA Received05/15/2023
Supplement Dates Manufacturer Received05/08/2023
06/05/2023
07/14/2023
Supplement Dates FDA Received05/17/2023
06/07/2023
07/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
EA DELTA CER INSERT 36IDX52OD; UNK HIP BONE SCREW; UNK HIP BONE SCREW; UNK HIP FEMORAL HEAD CERAMIC; UNK HIP FEMORAL STEM CORAIL; UNKNOWN HIP ACETABULAR CUP
Patient Outcome(s) Required Intervention;
Patient Age36 YR
Patient SexMale
Patient Weight55 KG
-
-