Model Number M00562670 |
Device Problems
Failure to Deliver Energy (1211); Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Event Description
|
It was reported to boston scientific corporation that a sensation short throw was used during a polypectomy procedure performed on (b)(6) 2023.During the procedure, after the sensation snare was inserted into the scope and connected to the generator, physician tried to activate the energy to cut the polyp; however, there was no response or heat, and the snare could not cut the polyp.The procedure was completed with another sensation snare.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
|
Manufacturer Narrative
|
Imdrf device code a090402 captures the reportable event of snare unable to deliver energy.Imdrf device code a050702 captures the reportable event of snare failure to cut the polyp.
|
|
Manufacturer Narrative
|
Imdrf device code (b)(4) captures the reportable event of snare unable to deliver energy.Imdrf device code (b)(4) captures the reportable event of snare failure to cut the polyp.Block h10: investigation results a captivator medium hexagonal stiff snare was received for analysis.Visual inspection of the returned device revealed no damages were found.The 2 in 1 connector passed dimensional testing.Electrical testing was performed, and the device was found within specification.No other problems were noted.The reported complaint of device failure to deliver energy and failure to cut could not be confirmed, since the device passed the electrical test, and the device cannot be functionally evaluated with respect to anatomical/procedural factors encountered during the procedure.Based on the information available and the returned device analysis, the most probable root cause for the reported complaint is no problem detected since the reported complaint cannot be confirmed.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.Block h11: correction to block g1 (mfr site facility name and mfr site address 2) the initial mdr.
|
|
Event Description
|
It was reported to boston scientific corporation that a sensation short throw was used during a polypectomy procedure performed on (b)(6) 2023.During the procedure, after the sensation snare was inserted into the scope and connected to the generator, physician tried to activate the energy to cut the polyp; however, there was no response or heat, and the snare could not cut the polyp.The procedure was completed with another sensation snare.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
|