• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANTS (BUNDLED COST); PATIENT-FITTED TMJ RECONSTRUCTION PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANTS (BUNDLED COST); PATIENT-FITTED TMJ RECONSTRUCTION PROSTHESIS Back to Search Results
Catalog Number CHG010
Device Problems Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2023
Event Type  Injury  
Event Description
It was reported that during placement of the implant, it was noticed it did not fit properly.There will be a second surgery performed with a new implant.
 
Manufacturer Narrative
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
 
Manufacturer Narrative
The event can be confirmed as the surgeon sent an image that showed the left mandibular component was not implanted.Moreover, a new order for the left mandibular component was submitted.On the date of the surgery, the surgeon could not implant the left mandibular component due to the changes in the patient¿s anatomy (the fossa component was implanted).The surgeon took a new scan and ordered a new mandibular component.The surgeon reported that the unusual delay between the planning medical grade ct scan, which was taken in (b)(6) 2022 after the first stage surgical procedure, and the eventual surgical implanting date in (b)(6) 2023 contributed to this event.This patient metabolically creates much heterotopic bone, which may have altered the mandibular anatomy on the lateral surface of the ramus to the point where the mandibular component no longer fits.
 
Event Description
It was reported that during placement of the implant, it was noticed it did not fit properly.There will be a second surgery performed with a new implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ UNILATERAL IMPLANTS (BUNDLED COST)
Type of Device
PATIENT-FITTED TMJ RECONSTRUCTION PROSTHESIS
Manufacturer (Section D)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer (Section G)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer Contact
rylee koole
6059 king drive
ventura 93003
8056503391
MDR Report Key16931081
MDR Text Key315209051
Report Number0002031049-2023-00036
Device Sequence Number1
Product Code LZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberCHG010
Device Lot Number2206141017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-