• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CARDIAC RESUSCITATOR BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CARDIAC RESUSCITATOR BAND Back to Search Results
Model Number MODEL 100
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2022
Event Type  malfunction  
Event Description
During patient use, the lifeband band clip had broken off and preventing the hinged belt guard of the lifeband (lot #unknown) from snapping in place to the autopulse platform.The autopulse platform (b)(6) displayed user advisory "(ua)12" (lifeband not present) error message.The customer performed 20-30 manual cpr and rosc was achieved at the hospital.Back at the station, a new lifeband (lot #175571) was also noticed with the same issue while just snapping the band clip off and on.The lifeband used on a patient was disposed and the lifeband used at the station was returned for evaluation.No consequences or impact to the patient.Please see the following related mfr report: mfr 3010617000-2022-01952 for the autopulse platform (b)(6).Mfr 3010617000-2022-02013 for the lifeband (lot #unknown).
 
Manufacturer Narrative
The report of the autopulse lifeband (lot # 175571) will not clip into the autopulse platform was confirmed during the visual inspection.The probable cause for the reported complaint was the broken locking tabs on the hinged belt guards of the lifeband.The observed damage appeared to be an isolated instance of excessive force applied to the hinged belt guards of the lifeband.In addition, during the visual inspection, unrelated to the reported complaint, noticed the autopulse lifeband head end of the band clip was missing.The missing piece was located inside the platform driveshaft likely as a result of user mishandling.The broken piece of the band clip was removed to allow the placement of the lifeband on the platform.Functional testing could not be performed due to the broken band clip.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaints reported for the lifeband lot number 175571.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LIFEBAND
Type of Device
CARDIAC RESUSCITATOR BAND
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave
san jose, CA 95131
4085411030
MDR Report Key16933832
MDR Text Key315531707
Report Number3010617000-2022-01949
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001571
UDI-Public00849111001571
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0701-01
Device Lot Number175571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/17/2022
Initial Date FDA Received05/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-