• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in fayetteville, arkansas.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a s5 roller pump gave a motor control failure error message at boot up.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to solve it, computer board (hkr), motor control board (hms) were replaced.In addition, lcd/led panel with touch screen, ribbon cables, silicone gasket and cable speaker were replaced as per s5 system user manual since they were the old version.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The reported error message is mostly caused by an interruption of the connection between hkr and hms boards, meaning that the hms board did not get information from the hkr board.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2010.Based on the information available and livanova technical intervention, it cannot be ruled out that the reported event was due to an electrical failure of the hkr board or hms board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16937295
MDR Text Key315380828
Report Number9611109-2023-00225
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/18/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received06/19/2023
Supplement Dates FDA Received07/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-