Model Number 0998-00-0800-53 |
Device Problems
Failure to Zero (1683); Inadequate User Interface (2958)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that , the cardiosave intra-aortic balloon pump (iabp) unit is unable to zero due to touch screen.
|
|
Event Description
|
It was reported during a routine check, the cardiosave intra-aortic balloon pump (iabp) unit is unable to zero due to touch screen.There was no patent involved.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) stated that device received damage to the top display bezel.Hinges showed signs of irreparable damage.Top display bezel when closed was moved 1/16th of an inch forward instead of being fush with the lower bezel.Upper lcd had a line in the display.Fse replaced assembly,top level display.Performed a full function and calibration check.Device operating within manufacturer specifications.Unit returned to customer and cleared for clinical use.
|
|
Event Description
|
N/a.
|
|