• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE INDUSTRIES, LP - NORTHFIELD ENTRAFLOH2O FEEDING PUMP; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES, LP - NORTHFIELD ENTRAFLOH2O FEEDING PUMP; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number ENTRAFLOH2O
Device Problems Crack (1135); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2023
Event Type  malfunction  
Event Description
We have been experiencing a high failure rate of medline entrafloh2o feeding pumps, related to the pump's iv(intravenous) pole clamps being overtightened since the pumps are not holding on the iv poles securely.This causes the following problems: pumps that are not tightened properly can slip or fall, leading to damage to the pump, staff and/or patients.When staff are attempting to properly secure the pumps to the iv poles and they over tighten the iv pole clamp, the rear housing of the pumps crack, exposing the pump to liquid ingress and possible falling of the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRAFLOH2O FEEDING PUMP
Type of Device
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP - NORTHFIELD
MDR Report Key16940721
MDR Text Key315386637
Report NumberMW5117567
Device Sequence Number1
Product Code PIF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENTRAFLOH2O
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/15/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-