Model Number 1458Q/86 |
Device Problems
Failure to Advance (2524); Separation Failure (2547)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/26/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Event Description
|
It was reported that the patient presented for routine implant of the left ventricular lead.During the procedure the physician attempted to insert the guidewire into the lead lumen, however guidewire was unable to advance through the lead.It was also noted that it was difficult to remove the guidewire from the lead.The patient was unstable.Further information was requested but not received.
|
|
Manufacturer Narrative
|
The reported event of failure to advance guidewire through the lead was not confirmed.As received, a complete lead was returned in one piece.The guidewire that was used in the field was not returned with the lead.The lead was tested with guidewire and was able to be inserted from the connector pin and out to the distal tip and was able to be removed with no anomalies noted.Electrical tests did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies.
|
|
Event Description
|
It was reported that the patient presented for routine implant of the left ventricular lead.During the procedure the physician attempted to insert the guidewire into the lead lumen, however guidewire was unable to advance through the lead.It was also noted that it was difficult to remove the guidewire from the lead the procedure was completed with a replacement lead.The patient was unstable.
|
|
Search Alerts/Recalls
|