This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, and h10.Review of the most recent repair record determined the reported event was confirmed.The cart was not powering on.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to a poor connection at the fuse spring terminal-to-fuse is the root cause for the fuse drawer thermal issues during use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|