FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT Back to Search Results
Model Number OPT944
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Low Oxygen Saturation (2477)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for usewith the airvo 2 series of humidification devices.Nhf therapy and the airvo 2 device are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Fisher & paykel healthcare (f&p) has requested for the complaint device to be returned for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in saint-quentin reported that the tubing of an opt944 optiflow + adult nasal cannula was found torn during use.The healthcare facility also reported that the patient who was hospitalized in an intensive care unit desaturated and the cannula was then replaced.No further consequence of the patient was reported.
 
Event Description
A healthcare facility in (b)(6) reported that the tubing of an opt944 optiflow + adult nasal cannula was found torn during use.The healthcare facility also reported that the patient who was hospitalized in an intensive care unit desaturated and the cannula was then replaced.No further consequence of the patient was reported.
 
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices and is also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.Method: the complaint opt944 optiflow + adult nasal cannula was received at f&p in new zealand for investigation, where it was visually inspected.Further information including the sequence of events, patient's condition, device set up was requested from the healthcare facility, however, no further details were provided.Our investigation is based on our evaluation of the subject device, initial information provided by the healthcare facility, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the opt944 optiflow + adult nasal cannula revealed that the tubing was torn near the 3-way connector and several areas of the tube were found to be stretched.Conclusion: our investigation was unable to determine the exact cause of the observed damage to the opt944 optiflow + adult nasal cannula.Based on our knowledge of the product and our observation that the tubing was stretched, the tubing was likely subjected to excessive force.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The subject opt944 optiflow + adult nasal cannula would have met the required specifications.The device history record (dhrs) for the lot was reviewed.No nonconformances were noted on the dhr and it is confirmed the lot was manufactured according to specification and passed all the required control measures.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state: "do not crush or stretch tube, to prevent loss of therapy." "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key16942841
MDR Text Key315364959
Report Number9611451-2023-00451
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-