BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G138 |
Device Problems
Failure to Capture (1081); High impedance (1291); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/10/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this patient was upgraded from an implantable cardioverter defibrillator (icd) to a cardiac resynchronization therapy defibrillator (crt-d) which required the addition of a left ventricular (lv) lead.The implant procedure was successful and normal measurements were confirmed the following morning and the patient was discharged.Subsequent data from remote monitoring revealed suspected loss of capture on the lv channel.Therefore, the patient was brought back to the hospital for an x-ray and re-testing.Pacing impedance measurements were greater than 3000 ohms and x-ray revealed the lv lead was not fully inserted into the device header.The pocket was re-opened, and this lead was re-inserted into the header.The set screw was tightened, and successful lv lead to device connection was confirmed with a tug test.Acceptable lead measurements were obtained.During the procedure, damage to the df1 and high voltage (hv) distal and proximal pin connectors was observed on the right ventricular (rv) lead.The hv segments were bent at a right angle where the pin connectors connect to the device.Normal shock impedance measurements were confirmed.The physician repaired the distal hv insulation instead of replacing the lead as there was scar tissue and multiple leads in the vein.Shock impedance in right ventricular (rv) coil to can configuration was 100 ohms following the repair.The procedure was completed without further incident.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that this patient was upgraded from an implantable cardioverter defibrillator (icd) to a cardiac resynchronization therapy defibrillator (crt-d) which required the addition of a left ventricular (lv) lead.The implant procedure was successful and normal measurements were confirmed the following morning and the patient was discharged.Subsequent data from remote monitoring revealed suspected loss of capture on the lv channel.Therefore, the patient was brought back to the hospital for an x-ray and re-testing.Pacing impedance measurements were greater than 3000 ohms and x-ray revealed the lv lead was not fully inserted into the device header.The pocket was re-opened, and this lead was re-inserted into the header.The set screw was tightened, and successful lv lead to device connection was confirmed with a tug test.Acceptable lead measurements were obtained.During the procedure, damage to the df1 and high voltage (hv) distal and proximal pin connectors was observed on the right ventricular (rv) lead.The hv segments were bent at a right angle where the pin connectors connect to the device.Normal shock impedance measurements were confirmed.The physician repaired the distal hv insulation instead of replacing the lead as there was scar tissue and multiple leads in the vein.Shock impedance in right ventricular (rv) coil to can configuration was 100 ohms following the repair.The procedure was completed without further incident.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation noted evidence indicating difficulty may have been encountered fully inserting an is-1/df-1 lead into the header of this device resulting in the reported clinical observations.Please refer to the description for more information regarding the specific circumstances of this event.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Search Alerts/Recalls
|
|
|