• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A Back to Search Results
Model Number SL-2000M2095D
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2023
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: event 3: the pink dialyzer connection cap on arterial end of bloodline is fused shut or is so tight it cannot be opened.If the cap can be removed, it leaks.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Event 3: a review of the manufacturing records for sl-2000m2095d, lot # 30155010 was performed and indicated there were no quality issues during the manufacturing of this lot.Although the shipment was delivered, the supplier stated the sample could not be located.If the sample does become available, the complaint will be reopened for further evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethelehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethelehem PA 18018
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key16945383
MDR Text Key315371146
Report Number2521402-2023-00141
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046955348763
UDI-Public(01)04046955348763
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSL-2000M2095D
Device Catalogue NumberSL-2000M2095D
Device Lot Number30155010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-