• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number LIFEVEST WCD 4000 SYSTEM
Device Problem Charging Problem (2892)
Patient Problem Lethargy (2560)
Event Date 03/21/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery charger was completed.The reported problem (unable to charge a battery) was confirmed.Upon investigation, the charger was unable to charge/recognize a battery.The failure was isolated to an internally open y1 crystal oscillator.The root cause of the open y1 crystal oscillator was unable to be positively identified.No adverse event resulted from the damaged charger.
 
Event Description
A us distributor reported that a patient's battery charger was unable to charge/recognize a battery pack.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA
Manufacturer Contact
zoll mfg corporation
121 gamma drive
pittsburgh, PA 15238
MDR Report Key16945929
MDR Text Key315379593
Report Number3008642652-2023-04810
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIFEVEST WCD 4000 SYSTEM
Initial Date Manufacturer Received 04/28/2023
Initial Date FDA Received05/17/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-