Catalog Number SGC0702 |
Device Problems
Leak/Splash (1354); Unstable (1667)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report a leak in the rotating hemostatic valve and the rotating hemostatic valve being unable to fully close.It was reported that during preparation for a mitraclip procedure, the steerable guide catheter (sgc) rotating hemostatic valve (rhv) was leaking.When set to closed, a wire was still able to pass through.The sgc was replaced, and the procedure continued.There was no clinically significant delay and no patient involved.No additional information was provided.
|
|
Manufacturer Narrative
|
The returned device analysis was unable to confirm both the reported leak and unstable.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported issue.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on the information reviewed and return device analysis, a cause for the reported leak and unstable could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Event Description
|
N/a.
|
|