It was reported that on (b)(6) 2023, a 29mm epic mitral valve was chosen for implant.During procedure when descending the valve into the patient's anatomy, the plastic support valve holder broke into 2-3 pieces.One of the pieces fell into the left ventricle.The piece was removed without difficulty, and the valve was and to be successfully implanted.There were no adverse consequences for the patient.There was no clinically significant delay in the procedure and the patient remained hemodynamically stable throughout.The patient is stable and in conventional hospitalization.
|
An event valve holder being damaged and broken during the procedure was reported.The valve holder was returned to abbott, and the investigation confirmed that the spool and plug were reversed on the base, with the spool in the location of the plug and the plug in the location of the spool.How or when the spool and plug became reversed could not be conclusively determined; however, it is consistent with the spool and plug being replaced in the incorrect location after the valve holder was disassembled in the field.There was damage noted in the hex cavity of both the spool and the plug.How or when the damage occurred could not be conclusively determined; however, it is consistent with damage during the implant procedure.Even though this could not be confirmed, the pattern of damage in the hexagonal cavity of the holder appears consistent with the use of an alternative tool.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications at the time of commercialization, which includes inspection of the holder and inspection of the final valve assembly.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|