Product event summary: the 4fc12 sheath was returned and analyzed.Visual inspection of the shaft area was performed.The inspection identified a shaft kink/twist at approximately 2.43 inches from the tip.The performance test with sentinel blackbelt leak tester was performed.All the performance tests were in the acceptable range.The shaft, side tube, and valve were all leak-tight with no apparent issue.In conclusion, the reported clinical issue, cardiac tamponade, occurred during the procedure.There is no indication of a relationship of the adverse event to the performance or a malfunction of the product.The sheath did not fail the returned product inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the patient had a cardiac tamponade and a pericardiocentesis was required to treat the patient.The case was aborted, the patient was not under general anesthesia. the patient was hospitalized.No further patient complications have been reported as a result of this event.
|