• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM1272
Device Problems Migration or Expulsion of Device (1395); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that the patient presented in clinic for follow-up.Upon interrogation, it was found that the pacemaker exhibited increased capture threshold.Chest x-ray was performed and confirmed that the device had migrated with twiddler's syndrome.The pacemaker was explanted and replaced.Patient condition was stable.
 
Manufacturer Narrative
Reported event of inadequate capture was not confirmed.The device above elective replacement indicator (eri) level was returned in one piece for analysis.Longevity assessment, output verification, capture test, and inspection of header and connectors were normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16958071
MDR Text Key315527969
Report Number2017865-2023-20457
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509572
UDI-Public05414734509572
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberPM1272
Device Catalogue NumberPM1272
Device Lot NumberA000122308
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RV LEAD
Patient Outcome(s) Required Intervention;
Patient Age90 YR
Patient SexFemale
Patient Weight59 KG
-
-