Catalog Number MP2002C-0006
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd maxguard¿ multi-fuse extension set with needleless connector was missing the maxplus valve.The following information was provided by the initial reporter: "connection problem between the maxplus luer tip and the luer lock of the infusion line with the same lot number : one piece didn't contain maxplus valve and the other one, it was difficult to screw the luer-lock male of the iv set with maxplus valve.".
 
Manufacturer Narrative
H.6.Investigation summary: a mp2002c-0006 sample was not available for investigation; however the customer indicated the complaint sample was from lot 22049397.The feedback provided by the customer suggests that the maxplus had not been correctly assembled onto the extension set.As part of the feedback the customer provided a photograph; analysis of the photograph confirmed the customer's experience as the maxplus had separated from the extension set.The details of this feedback were forwarded to the manufacturing site for investigation.A review of the production records for lot 22049397 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.The root cause of the customer¿s experience could not be determined as the sample was not available for investigation.In this instance, without a sample it is not possible to determine whether a manufacturing defect could have caused or contributed to the customer¿s experience.A review of the customer feedback database indicates that complaints of this nature are rare and there is currently no trend for issues of this nature against the mp2002c-0006 product.
 
Event Description
It was reported that the bd maxguard¿ multi-fuse extension set with needleless connector was missing the maxplus valve.The following information was provided by the initial reporter: "connection problem between the maxplus luer tip and the luer lock of the infusion line with the same lot number : one piece didn't contain maxplus valve and the other one, it was difficiult to screw the luer-lock male of the iv set with maxplus valve.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD¿ MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16959486
MDR Text Key315860029
Report Number9616066-2023-00959
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMP2002C-0006
Device Lot Number22049397
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-