Model Number FINETUNER ECHO [FINETUNER 2] |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
A finetuner echo was returned to service and repair.The device was returned due to defect.
|
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Manufacturer Narrative
|
According the information received, a finetuner echo was sent to service _ repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.
|
|
Event Description
|
A finetuner echo was returned to service and repair.The device was returned due to defect.
|
|
Search Alerts/Recalls
|