MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NITINOL STONE RETRIEVAL DEVICE 1.9F (0.63MM) 120 CM 11MM DAKOTA BASKET W/ OPENSU; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NITINOL STONE RETRIEVAL DEVICE 1.9F (0.63MM) 120 CM 11MM DAKOTA BASKET W/ OPENSU; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number CE2797
Device Problem Difficult to Open or Close (2921)
Patient Problem Insufficient Information (4580)
Event Date 04/11/2023
Event Type  malfunction  
Event Description
Provider asked for a dakota basket to retrieve the kidney stone.Staff checked the package integrity and expiration date, and opened the sterile supply.Upon checking the functionality of the basket, staff immediately noted that the basket would not close properly to be inserted into the instrument for stone retrieval.The dakota basket was removed from the sterile field and a new dakota basket was opened and passed the function test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITINOL STONE RETRIEVAL DEVICE 1.9F (0.63MM) 120 CM 11MM DAKOTA BASKET W/ OPENSU
Type of Device
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
marlborough MA 01752
MDR Report Key16963247
MDR Text Key315753421
Report NumberMW5117723
Device Sequence Number1
Product Code FFL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCE2797
Device Catalogue NumberM0063905010
Device Lot Number31272601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/18/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight83 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-