It was reported that a patient underwent an idiopathic ventricular tachycardia procedure with a thermocool® smart touch® sf bi-directional navigation catheter for which biosense webster¿s product analysis lab (pal) identified a hole on the pebax.It was initially reported by the customer that the carto 3 system was displaying error 106: force sensor error when the catheter was connected to the patient interface unit.To troubleshoot the cable was replaced without resolution.The catheter was replaced, the issue was resolved, and the procedure was continued.The caller noted that the force vector disappeared and the force stayed at 3 grams no matter where it was.There was no patient consequence.On (b)(6) 2023, the bwi pal revealed that a visual inspection of the returned device found a reddish material and a hole in the pebax.These findings were reviewed and assessed as an mdr reportable malfunction since the integrity of the device has been compromised.
|
Device evaluation details: the thermocool® smart touch® sf bi-directional navigation catheter was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed reddish material and a hole in the pebax.The magnetic and force features were tested and no errors were observed.The force values and the vector were observed within specifications.No force issues were observed.The blood found inside the pebax area may contribute to the force issue.A manufacturing record evaluation was performed and no internal action was found during the review.The issue reported by the customer was confirmed.The instructions for use contain the following information that should be considered: to ensure accurate force readings, verify that the force reading is near zero when the catheter is not in contact with tissue.If the force reading is not near zero when the catheter is not in contact with tissue, perform zeroing.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|