Catalog Number 2C4702K
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2023
Event Type  malfunction  
Event Description
It was reported that the bladder of a small volume folfusor ruptured.The issue occurred during filling with 5-fluorouracil.The fluid projected everywhere under the hood and onto the operator's gloves and on part of their shirt towards the lower abdomen and arms.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: device manufacture date: the lot was manufactured from august 25, 2022, to august 27, 2022.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation which showed a ruptured bladder.The reported condition was verified.The cause of the condition could not be determined, however, a possible cause may be due to inherent variation in the material, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16968236
MDR Text Key315825580
Report Number1416980-2023-02431
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080154
UDI-Public(01)00085412080154
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4702K
Device Lot Number22H032
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-