Model Number CDHFA300Q
Device Problems Defective Device (2588); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Event Description
During the initial implant procedure, the left ventricular (lv) lead was unable to be connected to the header of the device.Upon investigation, it appeared that the seal in the lv port of the device was missing.The device was replaced, and the implant procedure was successfully completed.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of lead connection issue was confirmed.The lv septum was found to be damaged upon receipt.Septum material was observed inside the lv (is-4) set screw hex cavity, and the set screw was slightly stripped.The connection issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16974408
MDR Text Key315676005
Report Number2017865-2023-20644
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032850
UDI-Public05415067032850
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA300Q
Device Catalogue NumberCDHFA300Q
Device Lot NumberP000166508
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2023
Initial Date FDA Received05/22/2023
Supplement Dates Manufacturer Received06/30/2023
Supplement Dates FDA Received07/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-